NSF Certified Class II A2 Biological Safety Cabinet



Model  BSC-4FA2(4') BSC-3FA2-GL(3')
BSC-3FA2-NA(3')
BSC-4FA2-GL(4')
BSC-4FA2-NA(4')
BSC-6FA2-GL(6')
BSC-6FA2-NA(6')
External Size (W*D*H)mm  1383*775*2295 1087*775*2265 1383*775*2295 1873*775*2295
Internal Size (W*D*H)mm  1210*600*660 910*600*660 1210*600*660 1700*600*660
Work Surface Height  750mm
Max Opening  550mm (22’’) 520mm(20.5’’) 550mm (22’’) 550mm (22’’)
Tested Opening  Safety height 200mm (8'') Safety height 254mm (10’’) Safety height 254mm
(10’’)
Safety height 254mm(10’’)
Airflow Volume Inflow 462m3/h(272cfm) 441m3/h(259cfm) 587m3/h(346cfm) 824m3/h(485cfm)
Down flow 65%:863m3/h(507cfm) 60%:649m3/h(382cfm) 60%:863m3/h(507cfm) 60%:1212m3/h(713cfm)
Exhaust 35%:462m3/h(272cfm) 40%:441m3/h(259cfm) 40%:587m3/h(346cfm) 40%:824m3/h(485cfm)
Front Window  Motorized. Two-layer laminated toughened glass ≥5mm. Anti UV   
HEPA Filter  Two, 99.999% efficiency at 0.3μm, filter life indicator   
Filter Guard Type  Aluminum alloy frame   
Noise  NSF 49≤67 dB/EN 12469≤65 dB   
Illumination  ≥1000Lux   
Display  LCD Display   
Waterproof Socket  2, Total consumption:≤500W   
Tap  Water tap*1, Gas tap *1   
Ground Resistance  ≤ 0.10Ω   
Power Supply  AC220V±10%, 50/60Hz; 110V±10%, 60Hz, Full load Amps: 9A, BTU/Hr: 1689   
Consumption  400W 300W 400W 500W
Alarm  Abnormal airflow velocity; Filter replacement; Front window at unsafe height.   
DC Motor  One ECM motor, 110V& 220V acceptable, Speed adjustable, high efficiency and low power consumption   
Material  Working area: 304stainless steel/  Frame and Decorative Plate: cold-rolled steel with anti-bacteria power coating   
UV Lamp 30W 18W 30W 40W
Germicidal UV lamp, Emission of 253.7 nanometers for most efficient decontamination   
Illuminating Lamp 18W*2 10W*2 18W*2 22W*2
LED lamp   
Standard Accessory  Illuminating lamp, UV lamp *2, Base stand, SS water & gas taps, Drain valve,
Waterproof sockets, Stainless Steel I.V. Bar with 10 hooks   
Optional Accessory  Armrest, Airflow tester, Formalin fumigation sterilizer, Infrared sterilizer,
Electric height adjustable base stand, Remote control   
Package Size (W*D*H)mm  1540*1080*1900 1250*1080*1840 1540*1080*1900 2020*1080*1900
Gross Weight(kg) 334 270 334 363

Three protection: operator, sample and environment. Airflow system: 70% air recirculation, 30% air exhaust A2 Cabinet is suitable for working with microbiological research in the absence of volatile or toxic chemicals and radionuclide.
 
ADVANTAGE
 
- Motorized front window.
- Large LCD display, all information displayed.
- Automatic air speed adjustable with filter block.
- With memory function in case of power-failure.
- Side & back wall is made up of single piece stainless steel.
- Interlock function: UV lamp and front window; UV lamp and blower, fluorescent lamp; Blower and front window.
- Front 10° slanted to offer operator comfort while working for long time, reduce glare and maximize reach into the work area.


Sorry, Currently No Video Uploaded.
Sorry, Currently No File Uploaded.

NSF Certified Class II A2 Biological Safety Cabinet


Please Leave Your Product Enquiry Here. We will Get Back To You Soon.

Product Specifications

  • Product SKU : -
  • Weight (GRAM) : -
  • Dimension : -